Evidence for Therapeutic Use of Ultrasound
(This material was borrowed from www.electrotherapy.org)
A .PDF FILE IS AVAILABLE WITH THIS MATERIAL AND ALL
DIAGRAMS - GO TO THE
DOWNLOAD PAGE
Ultrasound (US) is a form of MECHANICAL energy, not electrical energy and
therefore strictly speaking, not really electrotherapy at all. Mechanical
vibration at increasing frequencies is known as sound energy. The normal human
sound range is from 16Hz to something approaching 15-20,000 Hz (in children and
young adults). Beyond this upper limit, the mechanical vibration is known as
ULTRASOUND. The frequencies used in therapy are typically between 1.0 and 3.0
MHz (1MHz = 1 million cycles per second).
Sound waves are LONGITUDINAL waves consisting of areas of COMPRESSION and
RAREFACTION. Particles of a material, when exposed to a sound wave will
oscillate about a fixed point rather than move with the wave itself. As the
energy within the sound wave is passed to the material, it will cause
oscillation of the particles of that material. Clearly any increase in the
molecular vibration in the tissue can result in heat generation, and ultrasound
can be used to produce thermal changes in the tissues, though current usage in
therapy does not focus on this phenomenon (Williams 1987, Baker et al 2001, ter
Haar 1999, Nussbaum 1997, Watson 2000). In addition to thermal changes, the
vibration of the tissues appears to have effects which are generally considered
to be non thermal in nature, though, as with other modalities (e.g. Pulsed
Shortwave) there must be a thermal component however small. As the US wave
passes through a material (the tissues), the energy levels within the wave will
diminish as energy is transferred to the material. The energy absorption and
attenuation characteristics of US waves have been documented for several types
of tissue (see absorption section below).
Ultrasound Waves :
FREQUENCY - the number of times a particle experiences a complete
compression/rarefaction cycle in 1 second. Typically 1 or 3 MHz.
WAVELENGTH - the distance between two equivalent points on the waveform in
the particular medium. In an ‘average tissue’ the wavelength @ 1MHz would be
1.5mm and @ 3 MHz would be 0.5 mm.
VELOCITY - the velocity at which the wave (disturbance) travels through
the medium. In a saline solution, the velocity of US is approximately 1500 m sec-1
compared with approximately 350 m sec-1 in air (sound waves can
travel more rapidly in a more dense medium). The velocity of US in most tissues
is thought to be similar to that in saline.
These three factors are related, but are not constant for all types of
tissue. Average figures are most commonly used to represent the passage of US in
the tissues. Typical US frequencies from therapeutic equipment are 1 and 3 MHz
though some machines produce additional frequencies (e.g. 0.75 and 1.5 MHz).
The mathematical representation of the relationship is V = F.l (where V =
velocity, F = frequency and l is the wavelength.
US
Waveform
The US beam is not uniform and changes in its nature with
distance from the transducer. The US beam nearest the treatment head is called
the NEAR field, the INTERFERENCE field or the Frenzel zone. The behaviour of the
US in this field is far from regular, with areas of significant interference.
The US energy in parts of this field can be many times greater than the output
set on the machine (possibly as much as 12 to 15 times greater). The size
(length) of the near field can be calculated using r2/ where r= the radius of
the transducer crystal and = the US wavelength according to the frequency
being used (0.5mm for 3MHz and 1.5mm for 1.0 MHz).The US beam is not uniform and
changes in its nature with distance from the transducer.
As an example, a crystal with a diameter of 25mm operating
at 1 MHz will have a near field/far field boundary at : Boundary = 12.5mm2/1.5mm
» 10cm thus the near field (with greatest interference) extends for
approximately 10 cm from the treatment head when using a large
treatment
head and 1 MHz US. When using higher frequency US, the boundary distance is even
greater. Beyond this boundary lies the Far Field or the Fraunhofer zone. The US
beam in this field is more uniform and gently divergent. The ‘hot spots’ noted
in the near filed are not significant. For the purposes of therapeutic
applications, the far field is effectively out of reach.
One quality indicator for US applicators (transducers) is a
value attributed to the Beam Nonuniformity Ratio (BNR). This gives an indication
of this near field interference. It describes numerically the ratio of the
intensity peaks to the mean intensity. For most applicators, the BNR would be
approximately 4 - 6 (i.e. that the peak intensity will be 4 or 6 times greater
than the mean intensity). Because of the nature of US, the theoretical best
value for the BNR is thought to be around 4.0 though some manufacturers claim to
have overcome this limit and effectively reduced the BNR of their generators to
1.0.
Ultrasound Transmission
All materials (tissues) will present an impedance to the passage of sound
waves. The specific impedance of a tissue will be determined by its density and
elasticity. In order for the maximal transmission of energy from one medium to
another, the impedance of the two media needs to be the same. Clearly in the
case of US passing from the generator to the tissues and then through the
different tissue types, this can not actually be achieved. The greater the
difference in impedance at a boundary, the greater the reflection that will
occur, and therefore, the smaller the amount of energy that will be transferred.
Examples of impedance values can be found in the literature e.g. Ward 1986.

The difference in impedance is greatest for the steel/air interface which
is the first one that the US has to overcome in order to reach to body. To
minimise this difference, a suitable coupling medium has to be utilised. If even
a small air gap exists between the transducer and the skin the proportion of US
which will be reflected approaches 99.998% which in effect means that there will
be no transmission.
The coupling media used in this context include water, various oils,
creams and gels. Ideally, the coupling medium should be fluid so as to fill all
available spaces, relatively viscous so that it stays in place (!!), have an
impedance appropriate to the media it connects, and should allow transmission of
US with minimal absorption, attenuation or disturbance. For a good discussion
regarding coupling media, see Casarotto et al 2004, Klucinec et al 2000,
Williams 1987 and Docker et al 1982. At the present time the gel based media
appear to be preferable to the oils and creams. Water is a good media and can be
used as an alternative but clearly it fails to meet the above criteria in terms
of its viscosity. There is no realistic (clinical) difference between the gels
in common clinical use (Poltawski and Watson 2006).
In addition to the reflection that occurs at a boundary due to differences
in impedance, there will also be some refraction if the wave does not strike the
boundary surface at 90. Essentially, the direction of the US beam through the
second medium will not be the same as its path through the original medium - its
pathway is angled. The critical angle for US at the skin interface appears to be
about 15. If the treatment head is at an angle of 15 or more to the plane of
the skin surface, the majority of the US beam will travel through the dermal
tissues (i.e. parallel to the skin surface) rather than penetrate the tissues as
would be expected.
Absorption and attenuation :
The absorption of US energy follows an exponential pattern
- i.e. more energy is absorbed in the superficial tissues than in the deep
tissues. In order for energy to have an effect it must be absorbed, and at some
point this must be considered in relation to the US dosages applied to achieve
certain effects. Because the absorption (penetration) is exponential, there is
(in theory) no point at which all the energy has been absorbed, but there is
certainly a point at which the US energy levels are not sufficient to produce a
therapeutic effect.
A representation of the exponential absorption is shown in
the adjacent diagram.
As the US beam penetrates further into the tissues, a
greater proportion of the energy will have been absorbed and therefore there is
less energy available to achieve therapeutic effects. The half value depth is
often quoted in relation to US and it represents the depth in the tissues at
which half the surface energy is available. The will be different for each
tissue and also for different US frequencies. The table below gives some
indication of typical (or average) half value depths for therapeutic ultrasound.
(after Hoogland 1995)
| |
1 MHz |
3 MHz |
|
Muscle |
9.0 mm |
3.0 mm |
|
Fat |
50.0 mm |
16.5 mm |
|
Tendon |
6.2 mm |
2.0 mm |
As it is difficult, if not impossible to know the thickness
of each of these layers in an individual patient, average half value depths are
employed for each frequency
3 MHz 2.0 cm 1
MHz 4.0 cm
These values (after Low & Reed) are not universally
accepted (see Ward 1986) and some current research (as yet unpublished) suggests
that in the clinical (real world) environment, they may be significantly lower.
|
Depth (cm) |
3 MHz |
1 MHz |
|
2 |
50% |
|
|
4 |
25% |
50% |
|
6 |
|
|
|
8 |
|
25% |
To achieve a particular US intensity at depth, we must take
account of the proportion of energy which has been absorbed by the tissues in
the more superficial layers. The table gives an approximate reduction in energy
levels with typical tissues at two commonly used frequencies. More detailed
information can be found on the
Ultrasound Dose Calculation pages
Ultrasound Absorption in the Tissues
As the penetration (or transmission) of US is not the same in
each tissue type, it is clear that some tissues are capable of greater
absorption of US than others. Generally, the tissues with the higher protein
content will absorb US to a greater extent, thus tissues with high water content
and low protein content absorb little of the US energy (e.g. blood and fat)
whilst those with a lower water content and a higher protein content will absorb
US far more efficiently. It has been suggested that tissues can therefore be
ranked according to their tissue absorption.

Although cartilage and bone are at the upper end of this scale, the
problems associated with wave reflection mean that the majority of US energy
striking the surface of either of these tissues is likely to be reflected. The
best absorbing tissues in terms of clinical practice are those with high
collagen content – LIGAMENT, TENDON, FASCIA, JOINT CAPSULE, SCAR TISSUE (Watson
2000, ter Haar 99, Nussbaum 1998, Frizzel & Dunn 1982)
The application of therapeutic US to tissues with a low energy absorption
capacity is less likely to be effective than the application of the energy into
a more highly absorbing material. Recent evidence of the ineffectiveness of such
an intervention can be found in Wilkin et al (2004) whilst application in tissue
that is a better absorber will, as expected, result in a more effective
intervention (e.g. Sparrow et al 2005, Leung et al 2004).
Pulsed Ultrasound
Most machines offer the facility for pulsed US output, and for many
clinicians, this is a preferable mode of treatment. Until recently, the pulse
duration (the time during which the machine is on) was almost exclusively 2ms (2
thousandths of a second) with a variable off period. Some
machines now offer a variable on time. Typical pulse formats are 1:1 and
1:4 though others are available. In 1:1 mode, the machine offers an output for
2ms followed by 2ms rest. In 1:4 mode, the 2ms output is followed by an 8ms rest
period. The effects of pulsed US are well documented and this type of output is
preferable especially in the treatment of the more acute lesions. Some machines
offer pulse parameters that do not appear to be supported from the literature
(e.g. 1:9; 1:20). The duty cycle (% of time during which the machine gives an
output) will be 50% for the 1:1 mode and 20% for the 1:4 mode. This is a
relevant factor in dosage calculations and further details are included in the
dose calculation support material.
Therapeutic Ultrasound & Tissue Healing
One of the therapeutic effects for which ultrasound has been used is in
relation to tissue healing. It is suggested that the application of US to
injured tissues will, amongst other things, speed the rate of healing & enhance
the quality of the repair. The following information is intended to provide a
summary of some of the essential research in this field together with some
possible mechanisms through which US treatments may achieve these changes. It is
not intended to be a complete explanation of these phenomena or a comprehensive
review of the current literature. It may, none the less, provide some useful
basic information for clinical application.
The therapeutic effects of US are generally divided into:
THERMAL & NON-THERMAL.
THERMAL:
In thermal mode, US will be most effective in heating the dense
collagenous tissues and will require a relatively high intensity, preferably in
continuous mode to achieve this effect.
Many papers have concentrated on the thermal effectiveness of ultrasound,
and much as it can be used effectively in this way when an appropriate dose is
selected (continuous mode >0.5 W cm-2), the focus of this paper will be on the
non thermal effects. Both Nussbaum (1998) and ter Haar (1999) have provided some
useful review material with regards the thermal effects of ultrasound.
Comparative studies on the thermal effects of ultrasound have been reported by
several authors (e.g. Draper et al 1993, 1995a,b) with some interesting, and
potentially useful results.
It is too simplistic to assume that with a particular treatment
application there will either be thermal or non thermal effects. It is almost
inevitable that both will occur, but it is furthermore reasonable to argue that
the dominant effect will be influenced by treatment parameters, especially the
mode of application i.e. pulsed or continuous. Baker et al (2001) have argued
the scientific basis for this issue coherently.
Lehmann (1982) suggests that the desirable effects of therapeutic heat can
be produced by US. It can be used to selectively raise the temperature of
particular tissues due to its mode of action. Among the more effectively heated
tissues are periosteum, collagenous tissues (ligament, tendon & fascia) &
fibrotic muscle (Dyson 1981). If the temperature of the damaged tissues is
raised to 40 45°C, then a hyperaemia will result, the effect of which will be
therapeutic. In addition, temperatures in this range are also thought to help in
initiating the resolution of chronic inflammatory states (Dyson & Suckling
1978). Most authorities currently attribute a greater importance to the non
thermal effects of U/S as a result of several investigative trials in the last
15 years or so.
NON-THERMAL:
The non thermal effects of US are now attributed primarily to a
combination of CAVITATION and ACOUSTIC STREAMING (te Haar 99, Baker et al 2001,
Williams 1987). There appears to be little by way of convincing evidence to
support the notion of MICROMASSAGE though it does sound rather appealing.
CAVITATION in its simplest sense relates to the formation of gas filled
voids within the tissues & body fluids. There are 2 types of cavitation STABLE &
UNSTABLE which have very different effects. STABLE CAVITATION does seem to occur
at therapeutic doses of US. This is the formation & growth of gas bubbles by
accumulation of dissolved gas in the medium. They take apx. 1000 cycles to reach
their maximum size. The `cavity' acts to enhance the acoustic streaming
phenomena (see below) & as such would appear to be beneficial. UNSTABLE
(TRANSIENT) CAVITATION is the formation of bubbles at the low pressure part of
the US cycle. These bubbles then collapse very quickly releasing a large amount
of energy which is detrimental to tissue viability. There is no evidence at
present to suggest that this phenomenon occurs at therapeutic levels if a good
technique is used. There are applications of US that deliberately employ the
unstable cavitation effect (High Intensity Focussed Ultrasound or HIFU) but it
is beyond the remit of this summary.

ACOUSTIC STREAMING is described as a small scale eddying of fluids near a
vibrating structure such as cell membranes & the surface of stable cavitation
gas bubble (Burns 1981, Dyson & Suckling 1978). This phenomenon is known to
affect diffusion rates & membrane permeability. Sodium ion permeability is
altered resulting in changes in the cell membrane potential. Calcium ion
transport is modified which in turn leads to an alteration in the enzyme control
mechanisms of various metabolic processes, especially concerning protein
synthesis & cellular secretions.
The result of the combined effects of stable cavitation and acoustic
streaming is that the cell membrane becomes ‘excited’ (up regulate), this
increasing the activity levels of the whole cell. The US energy acts as a
trigger for this process, but it is the increased cellular activity which is in
effect responsible for the therapeutic benefits of the modality (Watson 2000,
Dinno et al 1989, Leung et al 2004).
MICROMASSAGE is a mechanical effect which appears to have been attributed
less importance in recent years. In essence, the sound wave travelling through
the medium is claimed to cause molecules to vibrate, possibly enhancing tissue
fluid interchange & affecting tissue mobility. There is little, if any hard
evidence for this often cited principle.
Ultrasound Application in relation to Tissue Repair
The process of tissue repair is a complex series of cascaded, chemically
mediated events that lead to the production of scar tissue that constitutes an
effective material to restore the continuity of the damaged tissue. The process
is more complex than be described here, but there are several interesting recent
papers and reviews including (Wener & Grose 2003, Toumi & Best 2003, Watson
2003, 2006, Hill et al 2003, Neidlinger-Wilke et al 2002, Lorena et al 2002,
Latey 2001).
INFLAMMATION:
During the inflammatory phase, US has a stimulating effect on the mast
cells, platelets, white cells with phagocytic roles and the macrophages
(Nussbaum 1997, ter Haar 1999, Fyfe & Cahal 1982, Maxwell 1992). For example,
the application of ultrasound induces the degranulation of mast cells, causing
the release of arachidonic acid which itself is a precursor for the synthesis of
prostaglandins and leukotreine – which act as inflammatory mediators (Mortimer &
Dyson 1988, Nussbaum 1997, Leung et al 2004). By increasing the activity of
these cells, the overall influence of therapeutic US is certainly
pro-inflammatory rather than anti-inflammatory. The benefit of this mode of
action is not to ‘increase’ the inflammatory response as such (though if applied
with too greater intensity at this stage, it is a possible outcome (Ciccone et
al 1991), but rather to act as an ‘inflammatory opimiser’. The inflammatory
response is essential to the effective repair of tissue, and the more
efficiently the process can complete, the more effectively the tissue can
progress to the next phase (proliferation). Studies which have tried to
demonstrate the anti inflammatory effect of ultrasound have failed to do so (e.g.El
Hag et al 1985 Hashish 1986, 1988), and have suggested that US is ineffective.
It is effective at promoting the normality of the inflammatory events, and as
such has a therapeutic value in promoting the overall repair events (ter Haar
99). A further benefit is that the inflammatory chemically mediated events are
associated with stimulation of the next (proliferative) phase, and hence the
promotion of the inflammatory phase also acts as a a promoter of the
proliferative phase.
Employed at an appropriate treatment dose, with optimal treatment
parameters (intensity, pulsing and time), the benefit of US is to make as
efficient as possible to earliest repair phase, and thus have a promotional
effect on the whole healing cascade. For tissues in which there is an
inflammatory reaction, but in which there is no ‘repair’ to be achieved, the
benefit of ultrasound is to promote the normal resolution of the inflammatory
events, and hence resolve the ‘problem’ This will of course be most effectively
achieved in the tissues that preferentially absorb ultrasound – i.e. the dense
collagenous tissues.
PROLFERATION:
During
the proliferative phase (scar production) US also has a stimulative effect
(cellular up regulation), though the primary active targets are now the
fibroblasts, endothelial cells and myofibroblasts (Ramirez et al 1997, Mortimer
and Dyson 1988, Young & Dyson 1990, Young & Dyson 1990b, Nussbaum 1997, 1998,
Dyson & Smalley 1983, Maxwell 1992). These are all cells that are normally
active during scar production and US is therefore pro-proliferative in the same
way that it is pro-inflammatory – it does not change the normal proliferative
phase, but maximises its efficiency – producing the required scar tissue in an
optimal fashion. Harvey et al (1975) demonstrated that low dose pulsed
ultrasound increases protein synthesis and several research groups have
demonstrated enhanced fibroplasia and collagen synthesis (Enwemeka et al 1989,
1990, Turner et al 1989, Huys et al 1993, Ramirez et al 1997). Recent work has
identified the critical role of numerous growth factors in relation to tissue
repair, and some accumulating evidence has identified that therapeutic US has a
positive role to play in this context (e.g. Reher et al 1999)
REMODELLING:
During the remodelling phase of repair, the somewhat generic scar that is
produced in the initial stages is refined such that it adopts functional
characteristics of the tissue that it is repairing. A scar in ligament will
not ‘become’ ligament, but will behave more like a ligamentous tissue. This is
achieved by a number of processes, but mainly related to the orientation of
the collagen fibres in the developing scar (Culav et al 1999, Gomez et al
1991) and also to the change in collagen type, from predominantly Type III
collagen to a more dominant Type I collagen (Vanables 1989, Forrest 1983). The
remodelling process is certainly not a short duration phase – research has
shown that it can last for a year or more – yet it is an essential component
of quality repair (El Batouty et al 1986, ter Haar 1987)
The application of therapeutic ultrasound can influence the remodelling
of the scar tissue in that it appears to be capable of enhancing the
appropriate orientation of the newly formed collagen fibres and also to the
collagen profile change from mainly Type III to a more dominant Type I
construction, thus increasing tensile strength and enhancing scar mobility
(Nussbaum 1998, Wang 1998). Ultrasound applied to tissues enhances the
functional capacity of the scar tissues (Nussbaum 1998, Huys et al 1993). The
role of ultrasound in this phase may also have the capacity to influence
collagen fibre orientation as demonstrated in an elegant study by Byl et al
(1996), though their conclusions were quite reasonably somewhat tentative.
Summary :
The application of ultrasound
during the inflammatory, proliferative and repair phases is not of value
because it changes the normal sequence of events, but because it has the
capacity to stimulate or enhance these normal events and thus increase the
efficiency of the repair phases (ter Haar 99). It would appear that if a
tissue is repairing in a compromised or inhibited fashion, the application of
therapeutic ultrasound at an appropriate dose will enhance this activity. If
the tissue is healing ‘normally’, the application will, it would appear, speed
the process and thus enable the tissue to reach its endpoint faster than would
otherwise be the case. The effective application of ultrasound to achieve
these aims is dose dependent.
REFERENCES
(1983).
“Ultrasonic assessment of stress fractures.” BMJ 286: 1479.
Baker, K. G., V. J. Robertson, et al. (2001).
"A review of therapeutic ultrasound: biophysical effects." Phys Ther 81(7):
1351-8.
Burns (1981) Ultrasound in Physiotherapy Pub:
Bristol Royal Infirmary
Byl, N. N., L. Hill Toulouse, et al. (1996).
"Effects of ultrasound on the orientation of fibroblasts: an in-vitro study."
Eur-J-Phys-Med-Rehabil 6(6): 180-4.
Casarotto, R. A., J. C. Adamowski, et al.
(2004). "Coupling agents in therapeutic Ultrasound: Acoustic and thermal
behaviour." Arch Phys Med Rehabil 85: 162-165.
Ciccone, C., B. Leggin, et al. (1991). "Effects
of ultrasound and trolamine salicylate phonophoresis on delayed-onset muscle
soreness." Phys Ther 71: 666-.
Culav, E. M., C. H. Clark, et al. (1999).
"Connective tissues: Matrix composition and its relevance to physical therapy."
Physical Therapy 79(3): 308-319.
Dinno, M. (1989). "The significance of membrane
changes in the safe and effective use of therapeutic and diagnostic ultrasound."
Phys Med Biol 34: 1543-.
Dinno, M. A., L. A. Crum, et al. (1989). "The
effect of therapeutic ultrasound on electrophysiological parameters of frog
skin." Ultrasound in Med and Biol 15(5): 461-470.
Docker, M.
F., D. J. Foulkes, et al. (1982). “Ultrasound couplants for physiotherapy.”
Physiotherapy 68(4): 124-125.
Draper, D., S. Sunderland, et al. (1993). "A
comparison of temperature rise in human calf muscle following applications of
underwater and topical gel ultrasound." JOSPT 17: 247-251.
Draper, D. and e. al (1995). "Rate of
temperature increase in human muscle during 1Mhz and 3Mhz continuous
ultrasound." J Orthopaedic and Sports Physical Therapy 22(4): 142-150.
Draper, D. O. and M. D. Ricard (1995). "Rate of
temperature decay in human muscle following 3 MHz ultrasound: the stretching
window revealed." Journal of Athletic Training 30(4): 304-7.
Draper, D. O., S. Schulties, et al. (1995).
"Temperature changes in deep muscles of humans during ice and ultrasound
therapies: an in vivo study." Journal of Orthopaedic & Sports Physical Therapy
21(3): 153-7.
Dunn, F.
and L. A. Frizzell (1982). Bioeffects of ultrasound. Therapeutic Heat and Cold.
F. Lehmann J, Rehabilitation Medicine Library: 386-.
Dyson, M.
and J. Niinikoski (1982). “Stimulation of tissue repair by therapeutic
ultrasound.” Infections in Surgery: 37-44.
Dyson, M. and D. Smalley (1983). Effects of
ultrasound on wound contraction. Ultrasound Interactions in Biology & Medicine.
R. Millner, E. Rosenfeld and U. Cobet. New York, Plenum Press: 151-158.
Dyson, M.
and J. Suckling (1978). “Stimulation of tissue repair by ultrasound: A survey of
the mechanisms involved.” Physiotherapy 64(4): 105-108.
El-Batouty, M. F., M. El-Gindy, et al. (1986).
"Comparative evaluation of the effects of ultrasonic and ultraviolet irradiation
on tissue regeneration." Scand J Rheumatology 15: 381-386.
El Hag, M., K. Coghlan, et al. (1985). "The
anti-inflammatory effect of dexamethazone and therapeutic ultrasound in oral
surgery." Br J Oral & Maxillofacial Surgery 23: 17-23.
Enwemeka,
C. S. (1989). “The effects of therapeutic ultrasound on tendon healing.” Am J
Phys Med Rehabil 68(6): 283-287.
Enwemeka, C. S., O. Rodriguez, et al. (1990).
"The biomechanical effects of low-intensity ultrasound on healing tendons."
Ultrasound Med Biol 16(8): 801-7.
Forrest, L. (1983). "Current concepts in soft
connective tissue wound healing." Br J Surgery 70: 133-140.
Frizzell, L. A. and F. Dunn (1982). Biophysics
of ultrasound. Therapeutic Heat and Cold. J. Lehmann. Baltimore, Williams &
Wilkins.
Fyfe, M.
C. and L. A. Chahl (1982). “Mast cell degranulation: A possible mechanism of
action of therapeutic ultrasound (Abstract).” Ultrasound in Med & Biol 8(Suppl
1): 62.
Gomez, M. A., S. L. Woo, et al. (1989).
"Medical collateral ligament healing subsequent to different treatment
regimens." J. Appl. Physiol. 66(1): 245-252.
Harvey, W., M. Dyson, et al. (1975). "The
stimulation of protein synthesis in human fibroblasts by therapeutic
ultrasound." Rheumatology & Rehabilitation 14: 237-.
Hashish, I., H. K. Hai, et al. (1988).
"Reduction of postoperative pain and swelling by ultrasound treatment: a placebo
effect." Pain 33(3): 303-311.
Hashish, I., W. Harvey, et al. (1986).
"Anti-inflammatory effects of ultrasound: Evidence for a major placebo effect."
Br J Rheumatology 25(1): 77-81.
Hill, M., A. Wernig, et al. (2003). "Muscle
satellite (stem) cell actiation during local tissue injury and repair." J Anat
203: 89-99.
Hoogland, R. (1995). Ultrasound Therapy. Delft,
Enraf Nonius.
Huys, S., B. S. Gan, et al. (1993). "Comparison
of effects of early and late ultrasound treatment on tendon healing in the
chicken limb." J Hand Ther 6: 58-59.
Kitchen,
S. and C. Partridge (1997). “Ultrasound, shortwave diathermy and laser: a survey
to examine patterns of use in England.” Br J Therapy Rehabil 4(2): 75-78.
Kitchen,
S. S. and C. J. Partridge (1990). “A review of therapeutic ultrasound:
Part 1 -
Background and physiological effects
Part 2 -
The efficacy of ultrasound
Part 3 -
References.” Physiotherapy 76(10): 593-600.
Klucinec, B., M. Scheidler, et al. (2000).
"Effectiveness of wound care products in the transmission of acoustic energy."
Phys Ther 80(5): 469-76.
Latey, P. (2001). "Aspects of inflammation." J
Bodywork and Movement Therapies 5(2): 124-131.
Lehmann, J. (1982). Therapeutic Heat and Cold.
Baltimore, Williams & Wilkins.
Lehmann,
J. F. and B. J. de Lateur (1989). Ultrasound, shortwave, microwave, superficial
heat & cold in the treatment of pain. Textbook of Pain. P. D. Wall and R.
Melzack. Edinburgh, Churchill Livingstone: 932-941.
Leung, M. C., G. Y. Ng, et al. (2004). "Effect
of ultrasound on acute inflammation of transected medial collateral ligaments."
Arch Phys Med Rehabil 85: 963-966.
Lorena, D., K. Uchio, et al. (2002). "Nornal
scarring: importance of myofibroblasts." Wound Repair Regen 10(2): 86-92.
Maxwell,
L. (1992). “Therapeutic ultrasound: Its effects on the cellular & mollecular
mechanisms of inflammation and repair.” Physiotherapy 78(6): 421-426.
McDiarmid,
T., P. N. Burns, et al. (1985). “Ultrasound and the treatment of pressure
sores.” Physiotherapy 71(2): 66-70.
Neidlinger-Wilke, C., E. Grood, et al. (2002).
"Fibroblast orientation to stretch begins within three hours." Journal of
Orthopaedic Research 20: 953-956.
Mortimer, A. J. and M. Dyson (1988). "The
effect of therapeutic ultrasound on calcium uptake in fibroblasts." Ultrasound
in Med & Biol 14(6): 499-506.
Nussbaum,
E. L. (1997). “Ultrasound: to heat or not to heat - that is the question.”
Physical Therapy Reviews 2: 59-72.
Nussbaum, E. (1998). "The influence of
ultrasound on healing tissues." J Hand Ther 11(2): 140-7.
Ramirez, A., J. A. Schwane, et al. (1997). "The
effect of ultrasoundon collagen synthesis and fibroblast proliferation in
vitro." Med Sci Sports Exerc 29: 326-332.
Reher, P., N. Doan, et al. (1999). "Effect of
ultrasound on the production of IL-8, basic FGF and VEGF." Cytokine 11(6):
416-23.
Szego C. (1974)
The lysosome as a mediator of hormone action. Recent Progress in Hormone
Research 30;171-233
ter Haar,
G. (1987). Recent advances in therapeutic ultrasound. Ultrasound: Method
Applications, Biological Effects and Hazzard Potential. M. Repacholi, M.
Grandolfo and A. Rindi. New York, Plenum Press.
ter Haar, G.
(1999). "Therapeutic Ultrsound." Eur J Ultrasound 9: 3-9.
Toumi, H. and H. M. Best (2003). "The
inflammatory response: friend or enemy for muscle injury?" Br J Sports Med 37:
284-286.
Turner, S., E. Powell, et al. (1989). "The
effect of ultrasound on the healing of repaired cockrel tendon: is collagen
cross-linkage a factor?" J Hand Surg 14B: 428-433.
Vanable, J. (1989). Integumentary potentials
and wound healing. Electric Fields in Vertebrate Repair. R. Borgens. New York,
Alan Liss Inc: 171-224.
Wang, E. D. (1998). "Tendon repair." J Hand
Ther 11(2): 105-10.
Ward, A.
R. (1986). Electricity, Fields and Waves in Therapy. Marrickville, Australia,
Science Press.
Watson, T. (2000). "Masterclass. The role of
electrotherapy in contemporary physiotherapy practice." Manual Therapy 5(3):
132-41.
Watson, T.
(2002.
Ultrasound
dose calculations.
In
Touch 101; 14-17
Watson, T. (2003). "Soft Tissue Healing." In
Touch 104: 2-9.
Werner, S. and R. Grose (2003). "Regulation of
wound healing by growth factors and cytokines." Physiol. Rev 83: 835-870.
Wilkin, L. D., M. A. Merrick, et al. (2004).
"Influence of therapeutic ultrasound on skeletal muscle regeneration following
blunt contusion." Int. J. Sports Med. 25(1): 73-77.
Williams, R. (1987).
“Production and transmission of ultrasound.” Physiotherapy 73(3): 113-116.
Young, S.
R. and M. Dyson (1990). “The effect of therapeutic ultrasound on angiogenesis.”
Ultrasound Med Biol 16(3): 261-9.
Young, S.
R. and M. Dyson (1990). “Macrophage responsiveness to therapeutic ultrasound.”
Ultrasound Med Biol 16(8): 809-16.
THERAPEUTIC ULTRASOUND - CONTRAINDICATIONS AND PRECAUTIONS
CONTRAINDICATIONS :
Do not expose either the embryo or foetus to therapeutic levels of
ultrasound by treating over the uterus during pregnancy
Malignancy,
malignant tissue or tissue thought to be malignant
Tissues in which
bleeding is occurring or could reasonably be expected
Significant
vascular abnormalities including deep vein thrombosis, emboli and severe
arteriosclerosis / atherosclerosis
Anaesthetic areas
(only really relevant for thermal applications)
Haemophiliacs not
covered by factor replacement
Application over :
The eye
The stellate ganglion
The cardiac area in advanced heart disease
Pacemaker or other active implanted device
The cranium
The gonads
Active epiphyses in children
PRECAUTIONS :
Always use the lowest intensity which produces a
therapeutic response
Ensure that the applicator is
moved throughout the treatment
Ensure that the patient is
aware of the nature of the treatment and its expected outcome
If a thermal dose is intended,
ensure that any contraindications that apply have been considered
Continuous ultrasound is
considered unwise over metal implants though low dose pulsed US is considered
acceptable
HAZARDS :
Reversible blood cell stasis can occur in small blood vessels if a standing
wave is produced while treating over a reflector such as an air/soft tissue
interface, soft tissue/bone or soft tissue/metal interface whilst using a
stationary applicator. Continuous movement of the treatment head removes this
hazard.
TREATMENT RECORD :
The operator should note :
Machine (model)
Machine settings – : frequency, intensity, time,
pulse parameters
Area treated
Any immediate or untoward effects
|